02271nas a2200337 4500000000100000008004100001260001300042653001600055653001100071653001300082653002200095653002200117653002300139653002000162653003900182653002500221653001700246653003200263653002300295653000900318653001600327100001400343700001600357700001600373245008100389300001100470490000700481050001600488520141500504022001401919 2011 d c2011 Jul10aAge Factors10aHumans10aHumidity10aMaterials Testing10aNeural Conduction10aObserver Variation10aPain Perception10aPeripheral Nervous System Diseases10aPhysical Stimulation10aRisk Factors10aSensitivity and Specificity10aSensory Thresholds10aSkin10aTemperature1 aHaloua MH1 aSierevelt I1 aTheuvenet W00aSemmes-weinstein monofilaments: influence of temperature, humidity, and age. a1191-60 v36 aHALOUA 20113 a

PURPOSE: To determine whether the buckling force of Semmes-Weinstein monofilaments is influenced by changes in temperature, humidity, and aging.

METHODS: We tested 16 Semmes-Weinstein monofilaments from North Coast Medical, varying in age from new to 12 years old. From each kit, we used the 2.83/0.07 g, 3.61/0.4 g, 4.31/2g, and 4.56/4 g filaments. Four observers tested the filaments under temperature/humidity conditions of 26°C/33%, 26°C/80%, 37°C/33%, and 37°C/80%.

RESULTS: Air temperature and humidity both significantly influence buckling force. At 26°C/33% the buckling force varies by an average of 2% from the standard data supplied by the manufacturer, whereas in the other 3 conditions (26°C/80%, 37°C/33%, and 37°C/80%), the buckling force can decrease by as much as 39%. Effect of age was small and not clinically relevant.

CONCLUSIONS: The significant changes in buckling force of the Semmes-Weinstein monofilaments in different temperature and humidity conditions might lead to a potential misinterpretation by the examiner regarding the level of peripheral nerve function loss. This could lead to incorrect diagnosis and treatment.

CLINICAL RELEVANCE: Examiners should be aware when using Semmes-Weinstein monofilaments that the environmental effects of temperature and humidity can affect the accuracy of measurement.